What is the accuracy of the HY9100 acidity meter standard


Regarding the accuracy of the HY9100 pH meter standard, it is difficult to give an exact figure because different manufacturers or different models of ph meters may have different accuracy specifications, and the specific accuracy may vary due to factors such as conditions of use, calibration status, and instrument aging. But in general, the accuracy of the pH meter is usually expressed in terms of measurement error or resolution.

In order to obtain the exact accuracy information of the HY9100 pH meter standard, it is recommended to refer to the following ways:

Information provided by the manufacturer: The manufacturer will usually provide a detailed product manual or technical manual, which will contain the accuracy specifications of the acidity meter. This is the most direct and accurate way to get accurate information.

Product label or nameplate: On the HY9100 acidity meter standard, there may be a label or nameplate with key information such as the type and accuracy of the instrument.

Professional websites or forums: On professional websites or forums in chemistry, environmental monitoring or related fields, you may find discussions or reviews about the HY9100 pH meter standard, which may contain information on accuracy. However, please note that this information may not be official data, so it needs to be treated with caution.

Contact the manufacturer or seller: If the exact accuracy information cannot be obtained by the above means, you can contact the manufacturer or seller of the HY9100 acidity meter standard directly, and they will usually provide detailed accuracy instructions and explanations.

It should be noted that even if the accuracy specification of the HY9100 acidity meter standard is known, it needs to be calibrated and maintained regularly during actual use to ensure the accuracy and reliability of its measurement results. In addition, in the use of acidity timing also need to pay attention to the correct operation method and the use of the environment, in order to avoid the measurement error caused by improper operation or environmental factors.

The correct use of HY9100 acidity meter measurement, you need to follow certain operating steps and precautions. Here is a detailed guide:

First, preparation

Calibration instrument:

Adjust the pH meter slope regulator to the appropriate position (usually 100%).

Adjust the temperature regulator to the corresponding temperature value according to the temperature of the solution being measured.

Two-point calibration using two buffer solutions closest to the pH of the sample to ensure measurement accuracy. The electrodes were inserted into each of the two buffer solutions, the beaker was shaken to make the solution even, and the pH was recorded after the reading stabilized. If the reading is unstable, adjust the positioning regulator until it is stable.

Cleaning electrode:

The glass electrode must be soaked in distilled water for more than one day and night before its first use. When usually used, it should also be soaked in distilled water.

Before use, clean the electrode with purified water and gently blot it with filter paper.

2. Measurement steps

Power on:

Turn on the power switch of the acidity meter and preheat it for a period of time (usually 20-30 minutes) to ensure the instrument is stable.

Insert electrode:

Insert the cleaned and drained electrode into the solution under test.

Ensure that the electrode bulb is fully immersed in the solution under test, and gently shake the beaker to make the solution even.

Read data:

Wait for the pH value displayed by the instrument to change within 1 minute, when the pH value displayed by the instrument is not more than ±0.05, the pH value displayed by the instrument is the pH value of the tested solution.

When recording the pH value of the measured solution, the temperature value of the measured solution should be recorded at the same time, because the pH value is affected by temperature.

Three, precautions

Electrode Maintenance:

The glass electrode should not be in contact with the strong water-absorbing solvent for too long, and should be operated as soon as possible when used in the strong alkali solution, and should be washed with water immediately after use.

When the glass film is stained with oil, alcohol should be used first, then carbon tetrachloride or ether, and finally soaked in alcohol, and then washed with distilled water.

The electrode can only be gently dried with filter paper after cleaning, do not wipe with fabric, so as to avoid static charge resulting in incorrect reading.

Measurement environment:

Avoid contact between the instrument and wet gas, the immersion of moisture will reduce the insulation of the instrument, and reduce its sensitivity, accuracy and stability.

During the measurement process, avoid severe vibration or impact on the electrode.

Continuous measurement:

For batches of liquid samples, the pH value can be continuously measured using the same pH meter. However, in the process of measurement, suspicious data need to be retested to confirm.

When retesting, the electrode needs to be slowly washed with distilled water 3-5 times, and then washed with the sample solution to be tested 3-5 times, and then inserted into the sample solution for measurement.

End of measurement:

After the measurement is completed, rinse the electrode with purified water, and then blot it with filter paper, and put the electrode protective sleeve (the sleeve is filled with the electrode protective liquid) for next use.

Following the above steps and precautions can ensure the measurement accuracy and stability of the HY9100 acidity meter, and provide reliable data support for scientific research and production.